2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 205
VE- 7.0.0 ISOLATE VERR
a) From debug screen 0, RECORD the valve numbers of the valves whose ‘dots’ are BLUE!
One of
these valve circuits may be the problem!
These procedures isolate between them.
b) Call debug screen 1. NOTE
VERR’s
value for reference later. If
VERR
EVER reaches 255 return to
(
page 196
) to reset it.
c) Watch
VERR
until if it increases above what was noted in part b OR for five (5) minutes whichever
comes first! TWO (2) possible scenarios:
1)
IF VERR
DOES NOT
increase above what was noted in part b:
A valve error is not
occurring at this time!
Turn Dialysate Flow on. If a Flow Error reoccurs proceed to
page 30,
2)
IF VERR increases above what was noted in part b:
VE- 7.0.1 ISOLATE VERR
a) At the distribution board (Figure below), unplug one of the RECORDED valves. This isolates the valves
solenoid and its wire harness.
Figure 40 – Distribution Board
b) Watch
VERR
until if it increases above what was noted in procedure number VE- 7.0.0, part b OR for
five (5) minutes whichever comes first! TWO (2) possible scenarios:
1)
IF VERR does DOES NOT increase:
The
last
valve unplugged OR
its blue wire harness is the
problem!
2)
IF VERR increases:
until all RECORDED valves have been
unplugged. If after all these valve(s) are unplugged AND if
VERR
is still increasing, THREE (3)
possible bad components, see the Component List below:
Component List
Swap in each, one at a time, returning to Dialysis Program in between, until
VERR
remains = 0 for
10 minutes!
1)
Actuator-Test Board;
2)
Actuator Cable;
3)
Distribution board
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...