2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 327
B)
FIGURE BELOW
, place one of the plugs, from the
FOUR-RESISTOR SET
into the Air
Sensor’s distribution board position i.e.
5
th
position
*
from LEFT
*
NOTE:
If CBE modified the resistor MUST into the CBE board pin for pin! ENSURE the top CBE
board pin is covered by the resistor!
C) If the resistor plug was placed properly, watching to thirty (30) seconds,
Chamber #69’s box MUST now say “No Air” always!
(page 328).
LEFT BLANK INTENTIONALLY