2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 365
CO- 7.3.4 ISOLATE NEGATIVE (-)12 VOLT DC SUPPLY
Per the Figure below, measure at the TEST Connector’s
-12V
(negative 12V) pin (pin 4, four pins from
left). TWO (2) possible scenarios:
1)
IF (and ONLY if) between -11.0 and -13.0 volts DC:
Swap in the listed components (see
COMPONENT LIST below) one at a time and in between return to (ABOVE) procedure number
(page 354) to test each component
COMPONENT LIST:
1)
Sensor Board;
2)
Actuator-Test Board;
3)
Functional Board (IC20?);
4)
Sensor Board cable;
5)
Distribution board.
2) IF NOT between -11.7 and -12.3 volts DC:
Proceed to
page 407
, procedure number
LEFT BLANK INTENTIONALLY