2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 435
FIL- 5.4.0 -250 NOT ACHIEVED OR IT DID NOT HOLD / ISOLATE FOUR-WAY
a)
Figure right
, clamp BOTH Four-Way dialyzer
Line tubing segments.
b) Can you achieve -250 mmHg and HOLD it (+/- 15
mmHg for one minute) now?
Yes
-250 achieved and HOLDS!
IMPORTANT!
Remove BOTH clamps from the Four-Way
Dialyzer Line segments and, leaving Dialysate Flow OFF, proceed to
page 466
, procedure
No CANNOT achieve or hold -250. Either the transducer protector at the meter is wet OR a Four-
Way tubing connection is leaking! See parts A and B below!
A) Locate and repair the leak.
B) Return to (ABOVE) procedure number
(
page 434
).
LEFT BLANK INTENTIONALLY