2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 401
EOS- 4.0.0, STEP #3 Isolate the Actuator Cable
a)
IMPORTANT! Turn the machine OFF to prevent damage!
b) Swap in (or check
2
) a known good actuator cable*. *To LOCATE the cable refer to
(page 9).
2
To check the actuator cable, which pump is issuing the EOS error?
IF ACID:
NOTE
that four (4) ACID PUMP connections (ONLY) will be checked and proceed to
page
SECTION 16 - CHECKING THE ACTUATOR BOARD CABLE
. If the actuator cable
checks good continue to procedure number EOS- 4.0.0, step #4.
IF BICARB:
NOTE
that four (4) BICARB PUMP connections (ONLY) will be checked and proceed to
page 521
SECTION 16 - CHECKING THE ACTUATOR BOARD CABLE
. If the
actuator cable checks good continue to procedure number EOS- 4.0.0, step #4.
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press ’Escape’ then call debug screen 0.
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch for pump symbol for up two (2)
minutes. Does it remain white now??
Yes
Remains white! The previous actuator cable is bad.
No Turns pink! See
EOS- 4.0.0, STEP #4
.
NOTE:
The previous actuator cable is probably
good!
EOS- 4.0.0, STEP #4:
If sure all above checks have been performed correctly, THREE (3) possible
bad components (see COMPONENT LIST below). Swap in, one at a time, testing the machine in between
to see if the new component fixed the problem:
COMPONENT LIST:
1)
Sensor Board
1
;
2)
Distribution board;
3)
Bad motherboard
1
To prevent a “Cond Offset Failure” place the machine into
T and C Mode.
Refer to
OPERATING
MODES
(page
Error! Bookmark not defined.
)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...