2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 241
T- 1.3.40 ISOLATE TEMPERATURE CONTROL
a) Figure right, place the
90°
C
(0.915 KΩ)
plug, from the
FOUR-RESISTOR
SET
into the
1
st
distribution board
position from the left, “CON-NTC”.
b)
Set your CALIBRATED volt meter to AC
voltage (~ V, V
AC
)!
c)
Turn the HEATER Switch ON!
d) Figure right, measure at the distribution board’s Heater
Connector, between the BLUE and BROWN wires. TWO
(2) possible scenarios:
1)
IF (and ONLY if) 100.0 volts AC or more:
Leaving
the 90°
C plug installed, proceed to
page 289
2)
IF LESS THAN 10.0 volts AC:
ENSURING the HEATER Switch was on AND the meter was set to
AC voltage
prior to measuring, see procedure number
(page 241).
T- 1.3.60 ISOLATE NTC #2
a) Call the Home screen to see the machine’s [
Temperature
] display.
b) Remove the 90
o
C plug from distribution board position, “CON-NTC”.
c) LEAVING “CON-NTC” VACANT, allow [
Temperature
] to increase to more than 38.0° C BEFORE
continuing to part d!
d)
Turn [Dialysate Flow] OFF (Dialysate Flow
on/off lamp blinks)!
e) Figure right, avoiding VACANT position #4, plug NTC
#2’s connector, into NTC #3’s position,
2
nd
position from
left (“MON-NTC”). Leave NTC #3 unplugged!
f) Call debug screen 4 to see
TEMP
(lower
left). TWO (2) possible scenarios:
1)
IF (and ONLY if) TEMP is more than 3.0:
(page 242).
2) IF TEMP is less than 3.0:
See parts A and B below:
A)
Replace NTC #2* with a known good. *To LOCATE NTC #2 refer to
B)
(page 242).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...