2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 78
F- 7.4.0 VERIFY DRAIN FLOW
a)
The “Select Program” screen remains up till instructed otherwise!
b) Was there ANY drain flow at all?
Yes
There was some flow!
Proceed to
page 79
, procedure number
No There was no flow at all! See procedure number
(page 78).
F- 7.4.2 NO DRAIN FLOW
Figure below, EXCEPT for valves #38 and #31, ENSURE are ALL OTHER valves, especially Valve #30,
are plugged PROPERLY into their distribution board positions?
Yes
Valves plugged in correctly! See procedure number
(page 79).
No
You located a valve plugged wrong! Return ALL VALVES to their distribution board PROPER
positions and return to
page 76
, procedure number
LEFT BLANK INTENTIONALLY