2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 252
E) Per the Calibration Booklet, perform TWO (2) calibrations:
Cal 1)
Temp Sensor AND
Cal 2)
Temp Control. After BOTH calibrations are complete continue to part F.
F)
Read before performing!
Return to
page 247
, procedure number T- 1.5.1 to see if the
calibrations fixed the problem. If (and ONLY IF) you return to part E continue to part G.
G) Turn the machine off and swap in a known good Sensor Board*. *To LOCATE the Sensor board
refer to
(page 9).
H) Perform THREE (3) calibrations
: Cal 1)
Temp Sensor AND
Cal 2)
Temp Control AND
Cal 3)
Cond Cells. After the calibrations are complete continue to part I.
I)
Read before performing!
Return to
page 247
, procedure number T- 1.5.1 to see if the new
board fixed the problem. If (and ONLY if) you return to part E continue to part J.
J)
Read before performing!
Turn the machine off and swap in a known good Functional Board.
*To LOCATE the Functional board refer to
(page 9).
K) Perform THREE (3) calibrations
: Cal 1)
Temp Sensor;
Cal 2)
Temp Control;
Cal 3)
Cond Cells.
After the calibrations are complete continue to L.
L) Return to
page 247
, procedure number T- 1.5.1 to see if the new board fixed the problem.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...