2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 322
Procedure # CO- 1.0.3 continued:
H) FIRMLY connect BOTH concentrate connectors to
JUGS
!!
I) Return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘CONFIRM’)
.
J) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
(page 322).
CO- 1.0.3.7 / CONDUCTIVITY LOW
A) Figure right, while watching through the inlet tubing,
disconnect the ACID connector from the jug, for seven (7)
seconds
THEN
FIRMLY reconnect!
B) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1)
IF (and ONLY if) air / acid is moving in one
direction ONLY, towards the machine!
Continue to part C.
2) IF air / acid is NOT moving OR moving back and
forth!
Proceed to
page 345
, procedure number
C) Figure right, watching through the inlet tubing,
disconnect the BICARB connector from the jug, for
seven (7) seconds
THEN
FIRMLY reconnect!
D) Watching for thirty (30) seconds, TWO (2) possible
scenarios:
1)
IF (and ONLY if) air / bicarb is moving, in one
direction ONLY, towards the machine!
See
page 323
).
2) IF air / bicarb is NOT moving OR moving back and forth!
Proceed to
page 345
, procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...