2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 498
SECTION 13 - DEAERATION PROBLEMS
DAIR- 1.0.0 ISOLATE ACTUATOR-TEST BOARD CONTROL
a) Enter
Service Mode → Calibrate Hydraulics → Deaeration Pressure but
DO
NOT
press ‘CONFIRM’
yet!
b) Per the Figure below, ENSURE the
[PUMP RATE]
box = 0!
Figure 83 – Calibrate Deaeration Pressure Screen
c) Figure right, is the
DEAERATION MOTOR’S
shaft rotating (Yes or No)?
Yes
Shaft rotating! If
[PUMP RATE]
DOES
NOT = 0 return to part a. If (and ONLY if)
[PUMP RATE]
= 0 replace the Actuator-
Test Board.
No NOT rotating! See procedure number
(page 498).
DAIR- 1.1.0 CHECK DEAERATION MOTOR
a)
ENSURE
the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it,
b)
SLAM*
the gauge into the Acetate/Acid rinse port.
*
ELSE pressure will not be read correctly!
c) Press ‘CONFIRM’. The screen says “
4. Adjust PUMP RATE…
” AND the
[Pump Rate]
box is yellow!
d) Does Loading Pressure (Rinse port gauge) PEAK to at least 16 psi?
Yes
16 psi or more! See procedure number
No Less than 16 psi! Perform parts a THROUGH d below:
a) If the
[Pump Rate]
window box is LESS THAN 210 set it to 210 and press ‘CONFIRM’. If
Loading Pressure remains less than 16 psi see part b. If at least 16 psi see procedure
number
(page 499).
b) Press ‘CONFIRM’ to save the calibration then turn the machine off.
Parts c and d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...