![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 154](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635154.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 150
e) Directing the tubing into cylinder, TURN THE WATER ON and measure for one (1) minute. TWO (2)
possible scenarios:
1)
IF (and ONLY if) 1000 ml or more collected:
(page 150).
2)
IF less than 1000 ml:
This is the problem! TWO (2) checks:
Check #1:
Figure right, if the (optional) incoming water filter is
present ENSURE it is clean!
Check #2:
ENSURE RO Product output is consistently more
than 1000 ml every minute! If not there is a problem
with the RO!
NW- 2.0.3 1000 ML OR MORE / ISOLATE PRESSURE REGULATOR #61
a) TURN THE WATER OFF and reattach the Incoming Water Tubing.
b) Figure right, loosen Inlet Pressure Regulator
#61’s Lock Nut.
c)
IMPORTANT!
TURN THE WATER ON!
d) Using METAL PLIERS, again clamp the tubing
between the gauge and the Heat Exchanger at the
location shown in
e) With the pliers REMAINING in place, adjust Inlet
Pressure Regulator #61’s center Adjustment Bolt
(Figure right) clockwise (inward) attempting to
increase pressure to more than 18 psi. TWO (2)
possible scenarios:
1)
IF (and ONLY if) more than 18 psi CAN be achieved
: Remove the pliers then allow one (1)
FULL minute. If a “No Water” alarm reoccurs return to (ABOVE) procedure number
(
page 143
). If the “No Water” alarm NEVER occurs there is no need to continue!
2)
IF 18 psi CANNOT be achieved:
Assuming incoming RO Product flow has remained more than
1000 ml every minute, Inlet Pressure Regulator #61 may be bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...