2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 226
Yes [
Conductivity
] window is white! See procedure number
(page 226).
No [
Conductivity
] window is red! Allow up to three (3) minutes OR until if the [
Conductivity
]
window turns white. TWO (2) possible scenarios below:
1)
IF the window turns white:
(page 226).
2) IF after three (3) minutes the window remains RED:
Proceed to
page 302
T- 1.0.133 VERIFY ‘OUT OF BYPASS’
Call debug screen 0. Figure right, is Valve #24’s ‘dot’ blue OR white?
1)
IF (and ONLY if) BLUE:
2) IF WHITE:
See parts a THROUGH c below:
a) Call debug screen 2. To ENSURE
the shunt door is CLOSED,
CVRCLS
(2
nd
column from left) = 1!
b) From the Home screen, ENSURE the [
Conductivity
] AND [
Temperature
] windows are pale
yellow/white!
c) Call debug screen 0. Watch Valve #24’s ‘dot’ for up to (2) more minutes OR until if it turns blue. TWO
(2) possible scenarios:
1)
IF (and ONLY if) the ‘dot’ turns blue
(page 226).
2)
IF the ‘dot’ remains white AND the [Temperature] window is white:
Proceed to
page 302
T- 1.0.134 VALVE #24’S ‘DOT’ IS BLUE / VERIFY ‘OUT OF BYPASS’ CIRCUIT
Call the Home screen. ENSURING [Dialysate Flow] is set to 500 ml/min, is the external flow indicator’s ‘bob’
rising at least ¼ way up in the sight tube?
Yes ‘Bob’ moving! Proceed to
page 232
, procedure number
No ‘Bob’ NOT moving! See parts a THROUGH c below:
a) ENSURE the shunt door is CLOSED!
b) ENSURE the [
Temperature
] AND [
Conductivity
] windows are NOT RED!
c) If ‘bob’ still is not moving AND Valve #24’s ‘dot’ is blue there may be masked Flow Error.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...