2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 165
CLEAN- 1.6.0 MORE THAN 300 ML
a)
Return the tubing to the DEAERATION
pump.
b) Figure right, return the Deaeration Pump to
distribution board position, “P20”!
c)
ENSURE the DEAERATION MOTOR
is running!
c) Turn Valve #65’s nut (Figure right) clockwise
(tighter) attempting to achieve a PEAK Loading
Pressure of between 22 and 25 psi?
Yes
Between 22 and 25 psi! Proceed to
page 171
, procedure number
No Less than 22 psi! See parts a AND b
below:
a) ENSURE more than 300 ml per minute
to the Deaeration pump i.e. no
recurring air lock!
b) Swap the following components (see
Component List below) one at a time,
with known good, attempting to adjust
Loading Pressure. When loading
pressure can be adjusted to more than
22 psi the last component swapped in
is the problem.
COMPONENT LIST: 1)
Deaeration Pump head;
2)
Deaeration Motor;
3)
Valve #65*;
4) Hydrochamber.
* Figure right, several threads should be visible
under Valve #65’s nut. If no threads are visible
either the wrong spring is installed** under Valve
#65’s nut OR Valve #65 is bad!
* * Refer to
(page 19).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...