2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 51
Scenario #3:
Referring to Figure C, for a ONE MINUTE PERIOD if EVER a
continuous weak stream that LASTS for five (5) seconds or
longer:
ENSURING the drain tube OPENING was pointed up, proceed to
page
Figure C
Scenario #4:
IF No Flow EVER:
If (and ONLY if) the Dialysate Flow on/off lamp is blinking
return to (ABOVE) procedure number
page 49
). If
NOT blinking proceed to
page 59
, procedure number
F- 6.1.0 PULSE – STOP – PULSE OR ‘DRIBBLE’ / ISOLATE FOR STUCK OPEN BC VALVE
A) Reattach the
drain tubing to the station!
B) ENSURING Dialysate Flow in ON (Flow on/off lamp is NOT blinking), ENSURE a good drain
connection by listening for or visually verifying flow.
C) If a TMP alarm is NOT present skip to part D. If a TMP alarm is present:
1)
Press and release the
‘Reset’ key then immediately press and hold it for three (3) seconds;
2)
Allow thirty (30) seconds;
3)
If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to part D.
D) Allow thirty (30) seconds BEFORE continuing to part E!
E) Call debug screen 4 to watch
PDIA
(left column) AND
ADIA
(right
column) for TWO (2) minutes. BOTH should REMAIN between
1.5 and 8.5 AND STABLE i.e. DOES NOT change more than 0.6
per minute!
F) TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA AND ADIA is STABLE between 1.5 and 8.5:
Proceed to
page 54
2) IF PDIA and / or ADIA
IS NOT
remaining between 1.5 and 8.5 and / or
IS NOT
stable:
See
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...