2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 49
F- 5.2.0 PREPARE TO ISOLATE VALVE #24 / VALVE #26
a)
IMPORTANT!
ENSURE the shunt door is CLOSED!
b) ENSURE Dialysate Flow is ON (Flow on/off lamp NOT blinking)!
c) Call the Home screen and ENSURE [Dialysate Flow] is set to 800 ml/min!
d) Figure right, is the [
Conductivity
] AND / OR the [
Temperature
]
window REMAINING ALWAYS RED?
Yes
REMAINING ALWAYS RED! See procedure number
(page 49).
No NOT REMAINING RED! See parts a THROUGH c below:
a) Figure right, inside the distribution board,
unplug the
5
th
connector cap
i.e. 6
th
position from the LEFT. This is Cond Cell
#7.
b)
ENSURE the [Conductivity]
window is now always red i.e.
Conductivity Cell #7 is unplugged!
c) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! Proceed to
page 111
, procedure number
No ‘Bob’ is NOT moving! See parts a AND b below:
a) Return Conductivity Cell #7’s connector to distribution board position “X7
COND”.
b)
Proceed to
page 50
, procedure number
F- 5.3.0 COND AND / OR TEMP WINDOWS ALWAYS RED / ISOLATE VALVE #24 / VALVE #26
Is the external flow indicator’s ‘bob’ rising at least ¼ way in the sight tube?
Yes
‘Bob’ moving! If SURE the
Conductivity
AND / OR
Temperature
window is ALWAYS
RED proceed to
page 111
, procedure number
No ‘Bob’ NOT moving! See procedure number
(page 50).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...