2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 189
CLEAN- 7.0.0 VALVE ERROR = 1 LONGER THAN TWO SECONDS
A)
To avoid damage turn the machine OFF!
B) Figure below, remove the distribution board cover.
C) FOUR (4) IMPORTANT checks!!
CHECK #1
ENSURE the ACTUATOR CABLE is plugged in
SECURELY
!
CHECK #2
Check the entire length of the ACTUATOR CABLE for damage!
CHECK #3
ENSURE all Valves are plugged PROPERLY into their distribution board positions!
CHECK #4
ENSURE the black PGND wire is plugged in and shows no signs of burning.
D) Was a problem located above?
Yes
Problem located! After the repair see procedure number
(page 189).
No problem located! Proceed to
page 190
, procedure number
CLEAN- 7.0.1 PROBLEM LOCATED
a)
Return to Heat Disinfect OR Rinse!
b) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
c) Watch the 2
nd
window down,
Valve Error
, for five (5) minutes. If it
DOES NOT
= 1 for longer than
two (2) seconds the repair solved the problem HOWEVER, if (and ONLY if) a Flow Error reoccurs see
(ABOVE) procedure number
page 157
).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...