2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 20
SECTION 1 – FLOW ERRORS IN DIALYSIS PROGRAM
A) Figure right, TWO (2) checks:
CHECK #1:
ENSURE the ‘To drain’ tubing is NOT
restricted AND (if used) the ‘quick connector’
is attached PROPERLY to the station!
CHECK #2:
ENSURE the Vent Tubing is NOT restricted!
NOTE:
Vent ‘slow dripping’ is normal with Flow Errors!
B) NEVER allow the DiaSafe
®
filter to hang from its tubing!
C) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish.
D)
Remove the ‘dummy chamber’ from the Level Detector
.
E) Call debug screen 2. Look at
! EMPTY
! (left column). TWO (2) possible scenarios:
1)
IF (and ONLY if) ! EMPTY = 1:
Continue to part F.
2)
IF ! EMPTY = 0:
Either the RED dialyzer line is NOT connected to the shunt properly OR FOUR
(4) possible bad components:
1)
Actuator-Test Board
OR;
2)
Functional Board
OR;
3)
Arterial
dialyzer line shunt door switch OR;
4)
Motherboard
F) ENSURE Dialysate flow is ON (Flow on/off lamp IS NOT blinking)!
G) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’!
H) ENSURE FIRM connections to Acid AND Bicarb JUGS!
I) Any active external hydraulic leaks seen?
Yes Active leak located!
Proceed to
page 553
No leaks! See part J.
J) WITHOUT LOOKING AWAY, Figure right, watch for a “No
Water” banner for one (1) minute! Does “No Water” EVER
appear, even if only once?
Yes
“No Water” appears! Proceed to
page 141
No
“No Water” NEVER
*
(page 21).
NOTE:
From here forward, as “No Water” alarms cause Flow Errors if (and ONLY if) a “No Water”
banner EVER appears address this alarm FIRST!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...