2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 218
T- 1.0.4 HEATER GOOD / ISOLATE TEMPERATURE SENSORS (NTCs’)
A)
BELOW
,
trace NTC #2 AND NTC #3’s cables from their distribution board connectors to the
sensors. TWO (2) IMPORTANT checks:
CHECK #1
ANY insulation damage indicates the sensor need to be replaced!
CHECK #2
ENSURE the sensors are not reverse connected with another component!
Figure 42 – Temperature Sensors (NTC)
B) During Checks #1 and #2, THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) no problems were located:
Proceed to
page 219
, procedure number
2)
IF (and ONLY if) insulation damage was located:
Replace the sensor(s) then see procedure
3)
IF a sensor was reverse connected:
AFTER correcting the problem see procedure number
(page 218).
T- 1.0.5 PROBLEM LOCATED DURING CHECKS 1 AND 2
Depending on where the problem ORIGINALLY occurred, TWO (2) possible scenarios:
1)
IF in Dialysis Program:
A)
Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM);
B)
Set [Dialysate Flow] to 800 ml/min;
C)
Allow six (6) minutes;
D)
If the temperature problem
(page 219).
2)
IF in Heat Disinfect:
A)
Return to Heat Disinfect;
B)
Allow up to thirty (30) minutes;
C)
If the problem
reoccurs put the machine in RINSE for fifteen (15) minutes then see procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...