2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 406
CR- 1.0.4 BETWEEN -11.0 AND -13.0 VOLTS / ISOLATE CELLS CALIBRATION
A) Turn the machine OFF.
B)
DO NOT allow the acid and LIQUID bicarb to run empty subsequently!
C) Place the machine into
Service Mode → Calibrate Sensors → Cond Cells.
D) ENSURING the external flow indicator’s ‘bob’ is moving up at least ¼ way up in the sight tube ALLOW
six (6) minutes BEFORE continuing to part E.
E) Set the external meter to measure Temperature (°C). If temperature is stable between 35.1 and 39.9°
C continue to part F. If (and ONLY if) temperature IS NOT between 35.1 and 39.9° C proceed to
page
F) Set the meter to measure Conductivity (mS).
G) Proceed per the screen’s instructions. Is the Cond Cells calibration successful?
Yes Calibration successful! See procedure number
(page 406).
No
Calibration IS NOT successful (“Operator Error”)! Swap in the listed component (see
Component List
below), one at a time, with known good then, in between, repeat
(page 406) to test each new component.
Component List
:
1)
Sensor Board;
2)
Test Board;
3
) Actuator-Test Board;
4)
Functional
Board’s EEPROM (IC20);
5)
Functional Board.
CR- 1.0.5 ISOLATE CONDUCTIVITY CIRCUIT
Turn the machine off then back on. Does the “Cond Ref Failure” OR “Cond Offset Failure” reoccur (Yes or
No)?
Yes
Problem reoccurs! Swap in the listed component (see the
Component List
below) one at a
time, with known good then, in between, repeat procedure number
(page 406) to test
the new component. When the Failure DOES NOT reoccur the last component swapped in is
the problem.
Component List
:
1)
Sensor Board;
2)
Test Board;
3
) Actuator-Test Board;
4)
Functional
Board’s EEPROM (IC20);
5)
Functional Board
No If the Sensor Board and / or Functional Board and / or Functional Board’s EEPROM (IC20) was
replaced it is necessary to perform ALL calibrations.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...