2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 32
F- 2.5.0 BC VALVES CYCLING OR FLWP NOT REMAINING = 255
a) The shunt door MUST remain open till instructed!
b) TWO (2) checks for an intermittent bad motor or Actuator Board:
CHECK #1:
a) Using the handle end
*
of a screwdriver, push on BOTH the FLOW AND DEAERATION
MOTOR shafts, release, then push again.
* Avoids potentially slicing your finger
b) Can you make EITHER motor stop rotating and REMAIN stopped (Yes or No)?
Yes
If ABSOLUTELY SURE one or both of the motors stops! THREE (3)
possible bad components:
1)
Bad Motor (probably brushes) OR;
2)
Bad
pump head OR;
3)
Bad Actuator-Test Board
1
.
1
To LOCATE the Actuator-Test board refer to
(page 9).
No Neither motor stops! See
CHECK #2
.
CHECK #2:
a)
IMPORTANT!
Turn Dialysate Flow OFF (Flow on/off lamp blinks).
b) Call debug screen 1. ENSURE
FLWP
(upper right)
=
255 i.e. Flow is
off!
c) The Deaeration Motor may be running but the Flow Motor should be OFF! Check
FLOW MOTOR rotation again!
d) Refer to the TABLE below:
F- 2.5.2 FLOW MOTOR NOT RUNNING / DEAERATION MOTOR RUNNING
a) Turn Dialysate Flow ON i.e. Flow on/off lamp NOT blinking!
b) From debug screen 1, ENSURE
FLWP
(upper right)
is LESS THAN 255
(i.e. Flow is on)!
(page 33).
Flow Motor
Rotating?
Deaeration Motor
Rotating?
Your Response
No
Yes
This is NORMAL! See procedure number
(page 32).
Yes
Yes or No
Flow Motor rotating! If
FLWP
is NOT = 255 you did not turn Flow
off! If
FLWP
= 255, TWO (2) possibilities:
1) E
ither the motors
are reverse connected at the distribution board OR
2)
The
Actuator-Test Board
1
is bad!
1
To LOCATE the board refer to
(page 9)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...