2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 492
AIR- 1.0.2 AIR SEEN AT L1 (OUTPUT FILTER #73)
TMP is between 300 and 340 mmHg:
Table 8 – Air Checks (1)
Location
Location Description
If air is seen go to:
If air is NOT seen the leak is from:
L2
Post Sample Port
L3
Dialysate Filter #73
L3
From Shunt (Figure 75)
L4
●
Equipped with Sample Port:
The
Sample Port is leaking
●
NOT Equipped with Sample Port:
Dialysate Filter #73 is leaking
L4
To Shunt (Figure 75)
L5
Dialyzer connector(s) or o-ring(s)
L5
Bottom of Flow Indicator #75
L6
Flow Indicator #75
L6
Input Valve #24 / Valve #26
L7
if DiaSafe
®
filter
equipped,
L9
if not
Valve #24 / Valve #26 mounting (L6) OR
Valves #24 / #26
L7
Input DiaSafe
®
Filter #90
L8
DiaSafe
®
Filter #90 OR Valve #24 / Valve
#26 mounting (
L6
) OR Valves #24 / #26
L8
Output Pre-Dialyzer Cond. Cell
L9
DiaSafe
®
Filter Test Valve (see Figure 82,
L9
Input Pre-Dialyzer Cond Cell #7
(Figure 75)
Procedure number
Pre-Dialyzer Cond. Cell #7
Figure 76 – Hydraulics Top View
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...