2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 284
T- 3.0.2 ISOLATE HEATER RESISTANCE
a)
Turn the HEATER Switch OFF!
b)
Set your volt meter to RESISTANCE
(Ω)!
c) Figure right, measure again at the Heater
Connector, between the BLUE and BROWN wires.
TWO (2) possible scenarios:
1)
IF (and ONLY if) less than 8.0
Ω OR more than
13.0
Ω:
ENSURING the HEATER Switch was OFF
before measuring, the heater is bad. To LOCATE
the heater, refer to
(page 129).
2) IF between 8.0 and 13.0
Ω:
See parts a THROUGH c below:
a)
Turn the HEATER Switch on!
b) To prevent issues with Pre Rinse, allow the
screen’s TOP
[Remaining Prerinse Time]
window =
0:00!
c) ENSURING the HEATER Switch is on AND
[Remaining Prerinse Time]
= 0:00, see
procedure
T- 3.0.3 ISOLATE RECIRCULATION
When
[Remaining Prerinse Time]
= 0:00 drain Valve #30 closes and Valve #29 opens resulting in no drain
flow. If Valve #30 is sticking open Heat Disinfect can take a long time to reach 80° C!
a) Figure right, if a ‘Quick Connector’ is present at
the end of the ‘to drain tubing’ an
adaptor
is
required!
b) Watching drain flow, allow forty five (45) seconds
THEN watch for thirty (30) seconds. TWO (2)
possible scenarios:
1)
IF (and ONLY if) drain flow HAS stopped:
This is normal! Reconnect the drain THEN see
2) IF drain flow did NOT stop:
This may or may not cause this temperature problem but must be fixed
before continuing! Perform parts a through c below:
a) Turn the machine OFF and swap in a known good, Valve #30*. *To LOCATE Valve #30 refer to
(page 193).
b) Place the machine into RINSE until it finishes i.e.
[Remaining Time]
= 0:00!
c) Return to Heat Disinfect! ENSURING
[Remaining Prerinse Time]
= 0:00 is there drain flow now
(Yes or No)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...