2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 350
CO- 5.0.0 UNSTABLE CONDUCTIVITY / CHECK PRESSURES
a) ENSURE Loading Pressure (Rinse port gauge) is cycling between its PEAK of somewhere
between 23 and 25 psi
AND
to NEVER LOWER than 11 psi.
b) Temperature influences Conductivity. If it changes more than +/- 0.2° C per minute address the
temperature problem first!
c) Turn the machine OFF to avoid air locking the deaeration pump!
d) The deaeration gauge is used next. ENSURE it reads
0 inHg before installing it!
e) Figure right, tee a gauge into the INLET tubing of the
Deaeration Pump.
f)
IMPORTANT!
Return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘CONFIRM’)!
g) Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration OKAY! Position the gauge so that
tubing in NOT pinched then see procedure
number
(page 350).
No NOTE this page number, as you will return here, THEN proceed to
page 498
CO- 5.1.0 CHECK FOR AIR LEAK
Per the Figure (above, right), using a flashlight, for ONE (1) MINUTE, WITHOUT LOOKING AWAY, check
for air bubbles flowing MOVING through the clear tubing at Conductivity Cell #7. Is air seen?
Yes
Air seen! Proceed to
page 495
, procedure number
No No air seen! See procedure number
(page 350).
CO- 5.3.0 NO AIR SEEN / ISOLATE PUMP ACCURACY
Call debug screen 0.Watch
Acid
versus
AAcid
(Figure right) for one (1) minute. The
values don’t matter but should remain /- 3 of each other?
Yes
AAcid = Acid!
(page 350).
No
AAcid
is NOT remaining
/- 3 of
Acid!
Proceed to
page 397
CO- 5.3.1 CHECK BICARBONATE STEPS
Watch
Bic
versus
ABic
(Figure right) for one minute. Remaining /- 3 of each
other?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...