2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 455
1)
IF (and ONLY if) WHITE:
Proceed to
page 456
, procedure number
2)
IF RED:
(page 455).
TMP- 2.0.10 TMP WINDOW RED
a) TMP is UNSTABLE if ANY one or more of the three (3) conditions below occur:
Condition #1:
Figure right,
if a
positive (‘+’) sign
*
appears in the TMP
window.
*
If a ‘+’ sign does NOT appear TMP is negative!
Condition #2:
If TMP does NOT remain between 0 and negative 450
mmHg (i.e. no ‘+’ sign)
Condition #3:
If TMP
changes more than
40 mmHg in one (1) minute
b) Is TMP UNSTABLE?
Yes
TMP
is unstable!
Proceed to
page 456
, procedure number
No
TMP
is STABLE! i.e. no ‘+’ sign AND TMP is between 0 and 450 AND is NOT changing more
(page 455).
TMP- 2.0.11 TMP WINDOW IS RED BUT TMP IS STABLE
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b) Allow thirty (30) seconds.
What color is the TMP window now? TWO (2) possible scenarios:
1)
IF (and ONLY if) WHITE:
(page 456).
2) IF remaining RED:
See parts a AND b below:
a) Per the Figure right, ENSURE
BOTH
CONDITIONS
below exist
:
Condition #1:
The Actual TMP bar is
remaining BETWEEN the
Upper (@ 520) and Lower
(@ 0) Alarm Limits!
Condition #2:
The TMP window is RED!
b) If (and ONLY if) BOTH conditions exist AND after multiple RESET attempts the TMP window
remains RED, FOUR (4) possible bad components:
1)
Actuator-Test Board OR;
2)
Sensor
Board
1
OR;
3
) Functional Board
1
OR;
4)
Motherboard.
1
To prevent “Cond Offset Failure” place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...