
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 446
TMP- 2.0.0 DEAERATION PRESSURE OKAY
a)
Turn the water OFF!
b) Figure right, ENSURE the external flow
indicator’s inner tapered tube is NOT
scored (i.e. tube really ‘cloudy’).
If
scored, replace it!
c) Allow a “No Water” alarm to occur
THEN, after one (1) minute, the
deaeration motor stops.
d) Remove the Deaeration gauge and
reattach the tubing!
e)
IMPORTANT! Turn the water on and allow the “No Water” alarm to go away!
f) Remove the hydraulic compartment from the cabinet!
g) Return BOTH concentrate connectors to their rinse ports.
h)
Place the machine into RINSE to pressurize the secondary hydraulic circuit!
From here forward, ENSURE a “No Water” alarm does not occur!
i)
Unless as external leak is immediate allow ninety (90) seconds minute BEFORE continuing to
procedure number
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...