2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 379
CO- 10.0.10 ISOLATE TEMP
a) Call debug screen 0. Figure right, ENSURE
Pre Dialyzer
Cond
(mS)
is within 0.4 of
TCD
.
b) ENSURE
Pre Dialyzer Temp
is between 35.5 and 38.5° C.
c) Is
Post Dialyzer Temp
within 1.2° C of
Pre Dialyzer Temp
?
Yes
Post Temp within 1.2 of Pre Temp! S
ee procedure
number
(page 379).
No Post Temp IS NOT within 1.2 of Pre Temp! Proceed to
page 380
, procedure number
CO- 10.0.11 POST TEMP WITHIN 1.2 OF PRE TEMP / ISOLATE POST DIALYZER COND
Post Dialyzer Cond
changes but overall is it remaining /- 0.4 mS of
Pre Dialyzer Cond
?
Yes
Post Cond within 0.4 of Pre Cond! See procedure number
(
page 385
No Post Cond IS NOT within 0.4 of Pre Cond! Proceed to
page 383
, procedure number
LEFT BLANK INTENTIONALLY