Figure 92 – Electronic Card Cage (Actuator-Test Board)
j)
Place the second meter on the matching numbered
PIN
on the solder (rear) side of the P2 connector.
Some components require checking multiple conductors
. 2.0
Ω
or less for ALL required
conductors?
Yes
The cable is good! Return it the distribution board. If referred to CHECKING THE
ACTUATOR BOARD CABLE, return to the procedure that brought you here.
No
Unplug the cable from the Actuator-Test Board’s P2 connector. Check inside the connector for
damaged
male
pins. If the pins are okay, replace the cable.
Table 11 – PUMPS Cable Plug / Actuator-Test Board P2 Pin Locations
PUMP
PLUG/PIN #
PLUG / PIN LOCATION
DEAERATION PUMP (1 of 2)
15
Bottom row, 8 pins from the rear of machine
DEAERATION PUMP (2 of 2)
16
Top row, 8 pins from the rear of machine
FLOW PUMP (1 of 2)
19
Bottom row, 10 pins from the rear of machine
FLOW PUMP (2 of 2)
20
Top row, 10 pins from the rear of machine
ACID PUMP (1 of 4)
1
Bottom row, first pin from the rear of machine
ACID PUMP (2 of 4)
2
Top row, first pin from the rear of machine
ACID PUMP (3 of 4)
3
Bottom row, 2 pins from the rear of machine
ACID PUMP (4 of 4)
4
Top row, 2 pins from the rear of machine
BICARB PUMP (1 of 4)
5
Bottom row, 3 pins from the rear of machine
BICARB PUMP (2 of 4)
6
Top row, 3 pins from the rear of machine
BICARB PUMP (3 of 4)
7
Bottom row, 4 pins from the rear of machine
BICARB PUMP (4 of 4)
8
Top row, 4 pins from the rear of machine
UF PUMP (1 of 4)
23
Bottom row, 12 pins from the rear of machine
UF PUMP (2 of 4)
24
Top row, 12 pins from the rear of machine
UF PUMP (3 of 4)
25
Bottom row, 13 pins from the rear of machine
UF PUMP (4 of 4)
26
Top row, 13 pins from the rear of machine
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...