2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 21
F- 1.0.1 NO WATER NEVER APPEARS / INITIAL LOADING PRESSURE CHECK
a) ENSURING the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it,
SLAM*
it into
the red Acetate/Acid rinse port. *OR ELSE pressure will not be read correctly!
b) Loading Pressure MAY OR MAY NOT be cycling but is it PEAKING to MORE than 15 psi?
Yes
More than 15 psi! Leaving the gauge installed, see procedure number
(page 21).
No Remains LESS THAN 15 psi! ENSURING the gauge was SLAMMED into the RINSE port, if
still less than 15 psi proceed to
page 116
, procedure number
F- 1.0.2 LOADING PRESSURE MORE THAN 15 PSI / ISOLATE “AIR IN DIALYSATE” FUNCTION
A)
Figure below
,
remove the Distribution Board cover.
B) The female Air Sensor’s connector is the
4
th
cap i.e. 5
th
position
FROM THE LEFT. If CBE
modified the connector plugs into the ‘CBE board’ two pins higher than the others!
C) Unplug the female Air Sensor’s connector.
DO NOT
remove the ‘CBE board’ under the connector!
D)
Leave the Air Sensor distribution board
position VACANT for now!
E) Call debug screen 0.
F)
Figure right
, reading the text box above Chamber
#69, TWO (2) possible scenarios (next page):
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...