2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 114
F- 14.0.6 BAD FLOW FROM VALVE #24 / ISOLATE VALVE #26
a)
Per the Figure below
, clamp Valve #26’s INPUT tubing!
Figure 24 – Valve #26
b)
Return to Calibrate Hydraulics → Flow Pressure.
c) Press ‘CONFIRM’ THEN ENSURE the Flow Motor is rotating!
d) Measure again from the blue dialyzer connector for one (1) minute.
e) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”. More than 800 ml collected (Yes or No)?
Yes
More than 800 ml collected! Return the connector to the door. Valve #26 is sticking open.
Return to procedure number
(page 112) to confirm this but if you return here:
TWO (2) possible bad components:
1)
Actuator-Test Board OR
2)
Valve #26.
No
Less that 800 ml collected! Valve #26 is okay! See procedure number
(page 115).
.