2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 303
STEP #3
a) The screen should say
“Connect a 1.255 K ohm resistor...”
b) Place the 80° C
(1.255 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-
Temperature Reference]
between 191 and 203?
Yes
Between 191 and 203! ‘Sharply press ‘CONFIRM’ then see
STEP #4
(page 303).
No NOT between 191 and 203! ENSURE the 80° C plug is placed properly at the
2
nd
position from
the left! If okay, see procedure number
STEP #4
a) The screen should say
“Connect a 0.915 K ohm resistor...”
.
b) Place the 90° C
(0.915 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-
Temperature Reference]
between 204 and 216?
Yes
Between 204 and 216! Save the calibration then proceed to
page 304
, procedure number
No NOT between 204 and 216! ENSURE the 90° C plug is placed properly at the
2
nd
position from
the left! If okay, see procedure number
(page 303).
T- 7.1.0 ERROR DURING TEMP SENSOR CALIBRATION
Perform this procedure if (and ONLY if) “Operator Error” OR “Actuator Board Error” banner occurred OR a
[Pre-Temperature Reference]
value was not in range.
a)
Turn the machine OFF!
b) Using a different
FOUR-RESISTOR SET
return to (ABOVE) procedure number
(page 302)
HOWEVER, if you return here see part c.
c)
Read before performing!
With the machine off, one at a time, swap in the listed components (see
COMPONENT LIST below), with known good then, in between, return to (ABOVE) procedure number
(page 302) to test each new component until the error banner does DOES NOT occur.
COMPONENT LIST: 1)
Actuator-Test Board
*
;
2)
Sensor Board
*
;
3)
Functional Board
*
;
4)
Sensor Board
cable;
5)
Distribution board;
6)
Motherboard
*
.
*
To LOCATE these boards refer to
(page 9)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...