2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 130
TROUBLESHOOTING MOTORS
MOTORS- 1.0.0 MOTOR TROUBLESHOOTING
A) The procedure that directed you here asked you to
NOTE
which motor (
Deaeration
OR
Flow
)
to
be
checked.
CHECK ONLY this motor!
B) Per the Figure below, the motor shaft is THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) NOT rotating at ALL:
Proceed to
page 133
, procedure number
2)
IF (and ONLY if) rotating CLOCKWISE (i.e. backwards!):
See procedure number
3)
IF rotating counterclockwise (CCW):
Proceed to
page 131
, procedure number
MOTORS- 2.0.0 MOTOR ROTATING CLOCKWISE (BACKWARDS)
a)
IMPORTANT!
Turn the machine OFF!
b) Per Figure right, remove the Torx screws (T-25) BUT
DO NOT
remove the cap to maintain the brushes in position!
c) Rotate the cap 180° then reinstall the screws and tighten them.
d) Return to the Program (Dialysis or Cleaning / Disinfection) that
originally brought you to TROUBLESHOOTING MOTORS!
e) Is the motor rotating counterclockwise (CCW) now?
Yes
Rotating CCW!
Place the
machine into Rinse for several minutes to eliminate a potential air lock.
No NOT rotating CCW! Return to (ABOVE) procedure number
(page 130).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...