2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 420
FIL- 2.0.6 ALARMS RESET / ISOLATE PRESSURE TEST
a) CLOSE the shunt door!
b) During this procedure you will look for air again through the INPUT TUBING at Chamber #69 but THIS
TIME while the Pressure Test’s “Get Neg TMP” AND “Neg Flow On” banners are up as this is when air
may be pulled in through a leak somewhere!
c) RESET all alarms!
d) Select the
[Test & Options]
tab at the bottom of the screen. TWO (2) possible scenarios:
1)
IF the [Pressure Test] button is BLUE: A)
Select [Pressure Test];
B)
Press ‘CONFIRM’;
C)
Allow the “Test: (Remove Air)” banner to expire THEN;
D)
IMMEDIATELY watch for air
bubbles at Chamber #69 for two (2) minutes!
2) IF the [Pressure Test] button is GRAY: A)
Select [Both
Tests];
B)
Press ‘CONFIRM’;
C)
Allow the Conductivity Test to
complete;
D)
Allow the “Test (Remove Air)” banner to expire
THEN;
E)
IMMEDIATELY watch for air bubbles at Chamber #69 for two (2) minutes!
e) Are air bubbles seen MOVING into Chamber #69?
Yes
Air seen! Proceed to
page 432
, procedure number
No air seen! Allow the tests to finish then one (1) more minute BEFORE performing parts A through
H below:
A) Call the Home screen. Set
[UF GOAL]
to 1000 ml; Set
[UF Time]
to 1:00 hr.
B) Press ‘CONFIRM’.
C)
RESET ALL alarms!
D) Turn the Blood Pump on so that it is rotating at more than 100 ml/min.
E) To avoid error, read parts F through H BEFORE performing them!
F) Press the front panel’s UF on/off key. The green lamp MUST stay
on solid!
G) When
TMP*
reaches 280 turn UF off IMMEDIATELY (i.e. UF lamp
OFF)!
*
If a “Filling Program” banner appears: 1
)
Repeat parts C) through G);
3)
If “Filling
Program” reoccurs more than twice proceed to
page 432
, procedure number
H) At Chamber #69’s INPUT tubing, IMMEDIATELY watch for air again. Air seen?
Yes
Air seen! Proceed to
page 432
, procedure number
No air seen! If you were prompted to Troubleshooting Filling Programs the original
problem may still be occurring OTHERWISE do NOT continue!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...