2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 11
NOTE B
Per the Figure below, machines produced prior to 2007 machines were equipped with separate
Actuator
and Test board. Machines produced in 2007 and after are equipped with an Actuator-Test board
which is a
single board:
•
The K/K2 Troubleshooting Guide, REV H, assumes all machines are equipped with the new style
Actuator-Test board i.e. REV H is NOT compatible with the old style Actuator board!
•
The Actuator-Test Board combines the functions of the
Old Style Actuator
and
Test Board
and uses
motherboard slot #3
(see
Figure 4A, page 9). To use the ‘combined’ feature the machine must be
equipped with Functional board software 3.29 or greater.
•
If an different Actuator-Test Board is swapped in and will be left in the machine a VOLTAGE
DETECTOR CALIBRATION (page 14) should be performed.
Figure 4C – Old / New Style Actuator Board