2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 474
SECTION 10 - PRESSURE TESTS FAILING
The
Pressure (Holding) Tests
(PHT) checks the hydraulic system for leaks which could compromise
patient fluid removal! These procedures troubleshoot ALL PHT failures:
A) Per
the Figure
below, the Automated Tests consist of two discrete tests:
1)
Alarms Tests
AND
2) Pressure Holding Tests
(Negative and Positive). If a Pressure Test fails one or more of the
following messages are listed in the Test & Options screen:
*Fail Remove Air, *Fail Get Neg TMP, *Fail Neg Stabilize, *Fail Neg Flow On, *Fail Get Pos TMP,
*Fail Pos Stabilize, *Fail Pos Flow Off”.
Figure 73 – Test & Options Screen
B)
Are the Tests CURRENTLY running i.e. screen says “ Test:…” OR “Testing”?
Yes Tests are running! See part C!
No
DO NOT start the tests!
Proceed to
page 476
, procedure number
C) Figure right, has the “Get Neg TMP” banner been presenting
for LONGER THAN ninety (90) seconds?
Yes “Get Neg TMP” lasting longer than 90 seconds! See procedure number
475).
No Allow the Tests to finish then
even if they pass
,
proceed to
page 476
, procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...