2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 495
AIR- 1.0.5 AIR SEEN FROM PRE-DIALYZER COND CELL #7
Figure below
,
ensuring
TMP is between 300 and 360 mmHg
, watch through the outlet (clear) tubing at
BOTH balancing chamber valves #31 AND #33. Is air coming from
1) ONE
OR
2)
BOTH
valves?
1)
IF ONLY ONE:
TWO (2) possibilities:
1)
Bad O-ring connection between the valve and the balancing
chamber;
2)
Cracked valve body.
2) IF BOTH:
Figure 79 – Balancing Chamber Valves #31 and #33
AIR- 1.0.6 CHECK FOR COMMON AIR LEAK
Are air bubbles MOVING INTO the machine through the clear acid and / or bicarbonate Inlet Tubing?
Yes
Air seen!
A)
Check the acid and / or the bicarbonates O-rings and plug;
B)
If using jugs ensure
they are full;
C)
Check the pickup wand(s) O-rings.
No air seen! See procedure number
(page 495).
AIR- 1.0.8 LEAK TEST HYDROCHAMBER
a) A procedure, in a different Section, is performed next.
NOTE
this page and procedure number (AIR-
1.0.8) as you may prompted to return to here!
b) BEFORE continuing to part c, proceed to
page 139
c) If a leak in the Hydrochamber was not located in part b return to Dialysis Program (
“Select Program” →
‘Dialysis’ → ‘CONFIRM’).
d) Return to (ABOVE) procedure number
page 490
) to continue looking for an air leak.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...