2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 250
T- 1.5.4 [TEMPERATURE] DOES NOT = 95° C BUT “TEMP OVER 95 DEGREES” REOCCURED
A)
Read this step before performing it!
Calibrate
Temp Sensor
per the Calibration
Procedures booklet but does “Operator Error”
OR “Actuator Board Error” occur during the
calibration?
Yes
An “Error” banner occurs! CAREFULLY repeat the
Temp Sensor
calibration but if
an
“Error” banner reoccurs proceed to
page 302
. If an Error banner
DOES NOT reoccur see part B.
No
An “Error” banner DOES NOT occur! See part B.
B)
Read before performing!
Return to
page 247
, procedure number
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 continue to part C.
C) BEFORE continuing to part D calibrate
Temp Control
per the Calibration Procedures booklet.
D)
Read before performing!
Return to
page 247
, procedure number
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 continue to part E.
E) BEFORE continuing to part F turn the machine off and swap in a known good Actuator-Test Board*. *To
LOCATE the board refer to
(page 9).
F) BEFORE continuing to part G, note this page number then proceed to
page 14
G)
Read before performing!
Return to
page 247
, procedure number
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 replace the Functional
Board* and perform all calibrations. *To LOCATE the board refer to
(page 9)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...