2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 408
d)
CAUTION!
Due to the time consuming work that may be performed if
the following resistance reading is misinterpreted proceed VERY carefully.
Do NOT read just
the meter’s numeric display, read ALSO its UNITS (“Ω”,
“KΩ”, “MΩ”) AND the location of its decimal point!
e) Measure resistance at TEST connector
pin 4
(4 pins from left). More than 75.0
Ω
(0.075 K
Ω)?
Yes
More than 75.0
No
Way less than 75.0 Ω. Possible -12 volt short. One of the card cage boards (which can be
isolated by removing them one at a time and measuring resistance until pin 4 is more than 75
Ω. After all boards have been removed and the measurement is still less than 75 Ω the
motherboard may be bad.
CR- 1.0.20 ISOLATE NEGATIVE TWELVE VOLT PROBLEM
Per Figure 4A and NOTE A (page 9), proceed according to how the machine is equipped (
1) Old
OR
2)
New Style Power Logic Board
:
1)
IF Old Style Power Logic Board:
Replace the Convertor
board
*
with a known good
or with a
known good New Style Power Logic Board and remove the Convertor board*
2) IF New Style Power Logic Board:
Replace it with a known good New Style Power Logic Board
1
OR
with a known good Old Style Power Logic Board AND Convertor board*
* CAUTION!
To avoid damage, the ‘Convertor board’ MUST be installed with the ‘solder side’ towards
the front of the machine AND matched pin for pin to the TEST connector.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...