
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 417
b) Figure below, remove the UF Pump’s output (TOP) tubing from the location shown!
c) Looking at the UF Pump’s vacant OUTPUT nozzle, TWO (2) possible scenarios:
1)
IF (and ONLY if) no fluid from the UF Pump:
Reattach the tubing then see procedure number
(page 418).
2) IF you see fluid output:
See parts a AND b below:
a) Call debug screen 0.
b) Per the Figure below, WITHOUT LOOKING AWAY, watch the UF Pump’s ‘dot’ for ONE (1)
FULL minute. Does it EVER BLINK between white and blue?
Yes
The UF ‘dot’ blinks! The Actuator-Test Board may be bad!
No
The UF ‘dot’ does NOT blink! Fluid from the UF Pump indicates bad or incorrectly
installed internal seals/springs OR the pump is upside down!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...