2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 185
CLEAN- 5.2.0 VERR = 0 / ISOLATE DIALYSATE SAMPLING OPTION
a) Call debug screen 0 to locate Valve #29’s and Valve #30’s
‘dots’ (Figure right). NORMALLY when one turns white the
other turns blue almost immediately.
b) WITHOUT LOOKING AWAY, watch BOTH ‘dots’ for three
(3) FULL minutes to see if they EVER
BOTH
turn white, at
the
SAME TIME
, for
LONGER THAN
thirty (30) seconds!
TWO (2) possible scenarios:
1)
IF (and ONLY if)
BOTH ‘dots’ turn WHITE, at the SAME TIME, for LONGER THAN thirty
(30) seconds!
(page 185).
2)
IF
BOTH dots,
at the SAME TIME, are
NEVER white LONGER THAN thirty (30) seconds!
Proceed to
page 186
, procedure number
CLEAN- 5.2.1 VALVE #30 AND VALVE #29 ‘DOTS’ REMAINING WHITE LONGER THAN 30 SECONDS
a) Place the machine into Service Program.
b)
From Service Program → Options → Hardware Options does the No
box, next to
Dialysate Sampling
(Figure right), have a blue ‘X’ it?
Yes
Dialysate Sampling
(page 185).
No
a) Move the ‘X’ to the No box THEN press ‘CONFIRM’. The ‘X’ MUST turn blue!
b) If (and ONLY if) the Flow Error was occurring in Heat Disinfect return to it. If not place the
machine in RINSE!
c) Watching for five (5) FULL minutes, TWO (2) possible scenarios:
1)
IF (and ONLY if) a Flow Error reoccurs:
2)
IF a Flow Error DOES NOT reoccur:
Problem solved! The
Dialysate Sampling
Hardware
Option was set to “Yes” accidently.
CLEAN- 5.2.2 ISOLATE FUNCTIONAL BOARD
When
the
Dialysate Sampling
Hardware Option = “No” then having Valve #30’s AND #29’s ‘dots’ white at
the same time is UNLIKELY and, if you made and error, could result in unnecessary work! See parts a
THROUGH c below!
a) Place the machine in RINSE.
b) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
c) WITHOUT LOOKING AWAY, watch Valve #30’s AND #29’s ‘dots’ again for four (4) FULL minutes. Do
they BOTH, at the SAME TIME, EVER turn WHITE again for more than thirty (30) seconds?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...