2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 653
DVF- 1.0.0 ISOLATE VERR / SPECIAL VALVE ERROR #38
a) ENSURING the machine is OFF, to avoid pulling cables loose, GENTLY open the card cage!
b)
IMPORTANT
! Set your CALIBRATED volt meter to RESISTANCE (Ω).
c)
Connect the meter’s black lead to chassis ground (see
(page 4)).
d) Figure below, at the top of the Actuator-Test Board locate its 50-pin P2 connector. ENSURE the
ACTUATOR ribbon cable is plugged in securely here!
If not this may be the problem!
e) Figure above, measure from the rear side of P2 at
pin #43
.
Should be more than 40.0
Ω
!
f) Figure right, reading the meter’s numeric AND
UNITS display, more than 40.0
Ω
?
Yes
MORE THAN 40
(page 653).
No
LESS THAN 40
Ω! Proceed to
page 660
, procedure number
DVF- 1.1.0 PIN 43 MORE THAN 40 OHMS
a) Close the card cage for now.
b)
IMPORTANT! Open the shunt door and LEAVE IT OPEN till instructed otherwise!
c) Turn the machine on AND return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
e) Allowing up to five (5) minutes, Figure right, does a
“Dial Valve Failure” OR “Act Byp Valve Fail” alarm
banner OCCUR?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...