2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 96
F- 9.0.27 40 PSI OR MORE / ISOLATE VALVE #78
a)
Turn the machine OFF!
b) Replace Valve #78* with a known good! *To LOCATE Valve #78 refer to the Figure previous page.
c) E
nter Service Mode → Calibrate Hydraulics → Flow Pressure.
d) Press ‘CONFIRM’ THEN ENSURE the FLOW MOTOR shaft is rotating!
e) Can you adjust the new Valve #78’s screw to a gauge pressure of between 35 and 36 psi (Yes or No)?
Yes
Pressure adjusts to between 35 and 36 psi! Figure
right, if Valve #78’s screw is turned so that no threads
are visible TWO (2) possible bad components:
1)
Bad
Flow Pump head OR;
2)
Bad Flow Pump motor
(possibly brushes).
No CANNOT adjust to between 35 and 36 psi! TWO (2)
possible bad components:
1)
Bad Flow Pump head OR;
2)
Bad Flow Pump motor (possibly brushes)
.
LEFT BLANK INTENTIONALLY