2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 450
c) Figure below, ENSURE the Air Sensor Connectors wires terminates with the brown wire to
the TOP male probe on top of Chamber #69 AND the blue wire to the bottom probe
!
d)
IMPORTANT!
Allow up to three (3) FULL minutes for
FILACT
to = 1. Does
FILACT
go
to 1?
Yes
FILACT
= 1!
Perform part A and B below:
A)
IMPORTANT!
Return the Air Sensor’s female connector PROPERLY
to its distribution board position! If CBE equipped, using a flashlight,
ENSURE the top CBE board pin is covered by the connector.
B)
Proceed to
page 452
, procedure number
No
FILACT
remains
= 0!
NOT LIKELY!
ENSURE the Air Sensor’s distribution board
position is COMPLETELY VACANT! After three (3) minutes, if (and ONLY if)
FILACT
remains = 0 see procedure number
(page 451).