2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 328
CO- 1.0.6 “NO AIR” ALWAYS / ISOLATE LOADING PRESSURE
a)
ENSURING
the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it
SLAM
*
it
into the Acetate/Acid rinse port.
*
or else pressure always read low
b) Loading Pressure may not cycle but is ‘OKAY’ if it achieves a peak of between 23 and 26 psi. TWO (2)
possible scenarios:
1)
IF (and ONLY if) pressure is ‘OKAY’:
Leaving the gauge installed, see procedure number
(page 328).
2) IF pressure is NOT ‘OKAY’:
Leaving the gauge installed, proceed to
page 27
procedure
number
F- 1.0.8
CO- 1.0.6.2 LOADING PRESSURE OKAY / ISOLATE INTERMITTENT FLOW ERROR
a) Call debug screen 0.
b) WITHOUT LOOKING AWAY, watch
Flow Error
for three (3) minutes
OR
until if it EVER = 1, even just
once, indicating a Flow Error:
1)
IF (and ONLY if) Flow Error is NEVER, EVER = 1!
Proceed to
page 329
, procedure number
NOTE:
If [
Conductivity
] has been ‘drifting’ between good and high watch
Flow Error
for up
to six (6) minutes to catch a potentially VERY intermittent Flow Error!
2) IF Flow Error = 1 even just once:
Proceed to
page 20
CO- 1.0.6.4 BC SWITCH NEVER = 897 PART 1
a) If previously removed, return the Air Sensor’s female connector to
5
th
distribution board
position
FROM THE LEFT. Figure below, if CBE modified it MUST plug into the CBE board pin for
pin.
(page 329).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...