2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 329
CO- 1.0.7 BC SWITCH NEVER = 897 PART 2
a) BE VERY SURE the blue connector is in KNOWN GOOD bicarb; the red connector in KNOWN
GOOD acid!!
b) Consider using the acid and bicarb jugs from another machine that has good Conductivity!
c) ENSURE [Dialysate Flow] is ON (Flow on/off lamp NOT blinking)!
d) From the Home screen, ENSURE [Dialysate Flow] is set to AT LEAST 500 ml/min!
e) If a problem was located in part a OR the acid AND / OR bicarb
was changed in part b allow six (6) minutes BEFORE continuing
to part e!!
f) Based on [
Conductivity
] now, THREE (3) possible scenarios:
Scenario #1:
IF (and ONLY if) LESS THAN 13.3 mS:
Proceed to
page 337
, procedure number
Scenario #2:
IF MORE THAN 14.3 mS:
(page 330)
Scenario #3:
IF (and ONLY if) STABLE between 13.3 and 14.3 mS:
Something may have
changed! If the original problem is no longer occurring DO NOT continue. If the
problem continues proceed to
page 335
, procedure number
LEFT BLANK INTENTIONALLY