2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 622
P- H.1.7 ISOLATE CARD CAGE
A)
IMPORTANT!
To prevent damage, turn the machine OFF!
B) Using ESD precautions, your choice, per the Board List below, swap in ONE of the card cage circuit
boards with known good, then continue to part C to test it.
BOARD LIST: 1)
Actuator-Test Board;
2)
Sensor Board
*
;
3)
Functional Board
*
.
* To prevent “Cond Offset Failure”, place the machine in
T and C Mode
. Refer to
OPERATING
MODES
(page
Error! Bookmark not defined.
)
C) Repeat the Voltage Detection Calibration. THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF “Operator Error” reoccurs:
Repeat parts A through C until “Operator Error” DOES NOT
occur indicating the last board replaced was the problem. If after all boards have been swapped
in AND if “Operator Error” still occurs the motherboard may be bad.
2)
IF 5V EST AND/OR 12V EST are not in range:
Repeat parts A through C until
5V EST AND / OR
12V EST
are in range indicating the last board replaced was the problem. If after all boards have
been swapped in AND the problem reoccurs the motherboard may be bad.
3)
IF “Operator Error” did NOT occur AND 5V EST AND 12V EST are within range:
The last
board swapped in was causing the problem!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...