2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 223
Yes
Between 64 and 76! ‘Sharply’ press ‘CONFIRM’ ONCE. If no Error banners* appear see
STEP
#2
(page 223).
*
If an “Operator Error” OR “Actuator
Board Error” banner appears see
procedure number
(page 223).
No
NOT between 64 and 76! ENSURE the 34° C plug is placed properly at the
2
nd
position from the
LEFT! If okay see procedure number
(page 223).
STEP #2
The screen should say
“Connect a 5.117 K ohm resistor...”
. Place the 41° C
(5.117 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-Temperature Reference]
between 157 and 169?
Yes
Between 157 and 169! ‘Sharply’ press ‘CONFIRM’ ONCE and see
STEP #3
(page 223).
No NOT between 157 and 169! ENSURE the 41° C plug is placed properly at the
2
nd
position from
the LEFT! If okay see procedure number
(page 223).
STEP #3
The screen should say
“Connect a 1.255 K ohm resistor...”
. Place the 80° C
(1.255 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-Temperature Reference]
between 191 and 203?
Yes
Between 191 and 203! ‘Sharply’ press ‘CONFIRM’ ONCE and see
STEP #4
(page 223).
No NOT between 191 and 203! ENSURE the 80° C plug is placed properly at the
2
nd
position from
the LEFT! If okay see procedure number
(page 223).
STEP #4
The screen should say
“Connect a 0.915 K ohm resistor...”
. Place the 90° C
(0.915 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-Temperature Reference]
between 204 and 216?
Yes
Between 204 and 216!
A)
Press ‘CONFIRM’ twice to SAVE the calibration;
B)
Proceed
page
, procedure number
No NOT between 204 and 216! ENSURE the 90° C plug is placed properly at the
2
nd
position from
the LEFT! If okay see procedure number
(page 223).
T- 1.0.10 ERROR DURING TEMP SENSOR CALIBRATION
A)
Read before performing!
Turn the
machine off! Using a different
FOUR-RESISTOR SET
return to
(page 222) HOWEVER, if you return here see part B.
B)
Read before performing!
Referring to COMPONENT LIST (below), swap in each, one at a time, with
known good and in between return to (ABOVE) procedure number
(page 222) to test each. If
you return here swap in the next component in the list until the error no longer occurs.
COMPONENT LIST: 1)
Actuator-Test Board
*
;
2)
Sensor Board
*
;
3)
Functional Board
*
; 4
)
Motherboard
*
*
To LOCATE these boards refer to
(page 9)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...