2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 227
T- 1.0.142 [TEMPERATURE] LESS THAN 35.1° C / ISOLATE HEATER CONTROL
a)
Turn the Heater Switch OFF!
b)
Set your CALIBRATED volt meter to AC volts (~ V, V
AC
)!
c) Allow
NINETY (90) FULL SECONDS BEFORE
continuing to part d!
d) Figure right, TIGHTLY HOLD the meter leads
BETWEEN the Heater Connector’s BROWN and BLUE
wires.
e) Watching the meter,
turn the Heater Switch on!
f) If EVEN JUST ONCE more than 100.0 volts is seen the
heater is on! TWO (2) possible scenarios:
1)
IF (and ONLY if) ALWAYS LESS THAN 100.0 volts AC:
The heater is staying off!
(page 227).
2)
IF EVER MORE THAN 100.0 volts AC:
Call debug screen 4 to see
TEMP CAL
(right column). More than one hundred and twenty (120)?
Yes
TEMP CAL
more than 120! See Proceed to
page 229
, procedure number
No
TEMP CAL
less than 120! Proceed to
page 302
, procedure number
T- 1.0.144 HEATER STAYING OFF
a) Figure right, inside the distribution board, UNPLUG the
1
st
connector from the left. This is NTC #2 and unplugging it
should turn the heater on full time!
b) Measure again BETWEEN the Heater Connector’s
BROWN and BLUE wires! TWO (2) possible
scenarios:
1)
IF (and ONLY if) more than 100.0 volts AC:
Turn the Heater Switch OFF
then see
(page 228
2)
IF (and ONLY if) ALWAYS less than 100.0 volts
AC:
Leaving NTC #2 unplugged till instructed,
proceed to
page 263
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...