![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 392](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635392.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 388
CO- 10.0.46 BOTH PRE AND POST COND ARE STABLE
a)
Pre Stbl Ct
(pre-conductivity stability counter) starts at 90 and decreases one unit
per second if
Pre Dev val
remains less than 135. If not already, after no more than
two (2) minutes, it should fall to 0!
b)
Post Stbl Ct
(post-conductivity stability counter) starts at 70 and decreases one
unit per second if
Post Dev val
remains less than 70. If not already, after no more
than two (2) minutes, it should fall to 0.
c) Call the Home screen and select the
[Kt/V AF]
button at the bottom of the screen.
d) Select the
[OLC Self-Test]
button then press
‘CONFIRM’. Does the “
Online Clearance
Self-Test”
banner appear?
Yes
The “
Online Clearance Self-Test”
banner appeared! IMMEDIATELY proceed to
page
, procedure number
No
The OLC Test did DID NOT start! ENSURING no alarms AND the shunt door is CLOSED,
TWO (2) possible scenarios:
1)
IF “Can’t Run OLC Test” banner appeared:
2)
IF
“
OLC steps not calculated” appeared:
Allow two (2) FULL minutes then attempt to
start the OLC Test again. TWO (2) possible scenarios i) or ii) below:
i)
IF “Online Clearance Self-Test” banner appears: I
MMEDIATELY
See
page 390
).
ii) IF “OLC steps not calculated” banner reoccurs:
Read before performing!
Proceed to
page 354
, procedure number
to calibrate COND
CELLS. If you return here (
“OLC steps not calculated”)
,
TWO (2) possible
bad components:
1)
Bad Actuator-Test Board;
2)
Bad Functional Board.
CO- 10.0.48 “CAN’T RUN OLC TEST” APPEARED
a) Reset all alarms!
b) Call debug screen 5.
c) If
Pre Stbl Ct
AND
Post Stbl Ct
are BOTH
= 0 OR goes to 0 within three (3) minutes proceed through
the following three (3) steps until if the OLC Test starts. If (and ONLY if) either one does NOT go to 0
within three (3) minutes proceed to
page 391
, procedure number
Step #1:
Allow two (2) minutes then attempt to start the OLC Self-Test again. If
“Online Clearance
Self-Test”
banner
appears see
(page 390). If
“Can’t Run
OLC Test”
reoccurs continue to step #2 next page.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...