2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 290
T- 4.0.1 PIN 6 MORE THAN 1.0 VOLTS DC / ISOLATE POWER LOGIC BOARD
a)
To prevent damage, turn the machine OFF!
b) Swap in a known good* Power Logic Board.
*
Known good = tested in another machine that does not allow temperature to exceed 39°.0 C!
c)
Set your volt meter to measure AC voltage (~ V, V
AC
).
d) Turn the machine on and return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
e) Figure right, measure at the distribution board’s
Heater Connector, between the BLUE and BROWN
wires. TWO (2) possible scenarios:
1)
IF (and ONLY if) more than 100.0 volts AC:
(page 290).
2) IF less than 10.0 volts AC:
Problem solved!
The
previous Power Logic Board is bad.
T- 4.0.2 ISOLATE THE SENSOR BOARD
NOTE:
The previous Power Logic Board is probably good!
a)
To prevent damage, turn the machine OFF!
b) Swap in a known good Sensor Board.
c)
Enter Service Mode → Options → Hardware Options.
d) Set
T and C Mode
to “Yes” and press ‘CONFIRM’ (the ‘X’ turns blue)
e) Turn the machine OFF!
f) Turn the machine on and return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
g) From the Home screen, ENSURING [
Temperature]
is
less than 40,0° C, again measure BETWEEN the
Heater Connector’s BLUE and BROWN wires. TWO (2) possible scenarios:
1)
IF (and ONLY if) more than 100.0 volts AC:
(page 291).
NOTE:
The previous Sensor Board is probably good.
2) IF less than 10.0 volts AC:
Problem solved!
The previous Sensor Board is bad. Perform parts A
through C below:
A) Return to Service Mode
→ Options → Hardware Options.
B) Set
T and C
mode to “No”.
C) Perform ALL calibrations.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...