2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 362
CO- 6.8.6 ISOLATE ACID PUMP
a) Figure right, check the Acid Connector’s O-rings
AND ENSURE the filter is clean AND the adaptor
plug is screwed in tight!
b) Read parts c and d before performing them.
c) Perform the ACID (CONCENTRATE) PUMP VOLUME CALIBRATION per the CALIBRATION
PROCEDURES booklet.
d) Are you able to accurately enter the burette
volumes and ‘CONFIRM’ them without an
“Operator Error” banner occurring (Figure right)?
Yes
“Operator Error” did NOT occur! See procedure number
(page 362).
No “Operator Error” occurred! Problem with the acid pump! See (ABOVE) procedure number
CO- 6.8.8 ISOLATE BICARBONATE PUMP
a) Figure right, check the Concentrate Connector’s O-
rings AND ENSURE the filter is clean AND the adaptor
plug is screwed in tight!
b) Perform the BICARBONATE PUMP VOLUME
CALIBRATION per the CALIBRATION PROCEDURES
manual
!
c) Are you able to enter the burette volumes and
‘’CONFIRM’ them without an “Operator Error”
banner occurring (Yes of No)?
Yes
“Operator Error” did NOT occur! Repeat the Cond Cells calibration. If the external meter’
conductivity is NOW between 13.0 and 14.4 return to ABOVE procedure number
(page 356). If NOT, assuming you performed all procedures correctly, the Troubleshooting
Guide cannot locate the problem.
No “Operator Error” occurred! Problem with the bicarb pump! See (ABOVE) procedure number
(page 349).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...